Output wavelength (source) |
Port 1: 1310, 1550 nm Laser APC / Port 3: 850, 1300 nm LED APC |
Output power (source) |
-3 dBm ±1 dB (@ 9/125 µm for 1310, 1550 m) / -23 dBm ±1 dB (@ 62.5/125 µm for 850, 1300 nm) |
Adjust output power (source) |
Adjustable over 7 dB in 0.01 dB steps (for 1310, 1550 nm) / Fixed (for 850, 1300 nm) |
Short term stability for 15 min, typical ±2 °C, after warm up, ORL < -25 dB (source) |
0.03 dB (for 1310, 1550 nm) / 0.01 dB (for 850, 1300 nm) |
Stability over temperature (source) |
0.2 dB (max for 1310, 1550 nm) / 0.35 dB (max for 850, 1300 nm) |
Wavelength initial tolerance @ 25 °C (source) |
20 nm (for 1310, 1550 nm) |
Wavelength drift (source) |
0.4 nm per °C (typical for 1310, 1550 nm) / 0.4 nm per °C (typical for 850, 1300 nm) |
Mode controlled (source) |
850, 1300 nm (50/125µm compliant: IEC 61280-4-1 {Ed.1.0}, TIA 526-14A & TIA TSB-178) |
Spectrum width @ FWHM (source) |
3 nm (typical for 1310, 1550 nm) |
Reconnection repeatability @ 95% confidence (source) |
0.1 dB (for 1310, 1550 nm Laser) / 0.05 dB (for 850, 1300 nm LED) |
Length range (2-way Autotest only) |
6 dB @ 62.5 µm (for 850, 1300 nm) / 30 dB (for 1310, 1550 nm) |
Length accuracy (2-way Autotest only) |
0.01% ± 4 meters |
Length resolutions |
1 meter (0.000~9.999 Km), 10 meters (10.00~99.99 Km), 100 meters (100.0~127.9 Km) |
Loss range (2-way Autotest) |
27 dB @ 62.5 µm (for 850, 1300 nm) / 47 dB (for 1310, 1550 nm) |
Loss repeatability & loss linearity (2-way Autotest) |
0.06 dB (for 850, 1300 nm) / 0.04 dB (for 1310, 1550 nm) |
Loss range (1-way Autotest) |
27 dB @ 62.5 µm (for 850, 1300 nm) / 47 dB (for 1310, 1550 nm) |
Loss repeatability & loss linearity (1-way Autotest) |
0.06 dB (for 850, 1300 nm) / 0.04 dB (for 1310, 1550 nm) |
ORL range |
0 ~ >60 dB (1310, 1550 nm) / 0 ~ 30 dB @ 62.5 µm |
ORL port isolation |
> 60 dB (1310, 1550 nm) / > 25 dB (850, 1300 nm) |
ORL resolution |
For 1310, 1550 nm: 0.01 dB (@ 0 ~ 50 dB), 0.1 dB (@ 50 ~ 65 dB) / For 850, 1300 nm: 0.01 dB (@ 0 ~ 30 dB), 0.1 dB (@ 30 ~ 45 dB) |
ORL accuracy |
0.2 dB |
|
|
Detector type (meter) |
InGaAs |
Response wavelength (meter) |
600 ~ 1700 nm |
Calibration wavelengths (meter) |
780, 820, 850, 980, 1270, 1290, 1300, 1310, 1330, 1350, 1370, 1390, 1410, 1430, 1450, 1470, 1490, 1510, 1530, 1550, 1570, 1590, 1610, 1625, 1650 nm |
Accuracy (meter) |
1 % (0.06 dB) |
Mid-range linearity (meter) |
0.04 dB (typical) |
Power range (meter) |
+10 ~ -60 dBm (780 ~ 980, 1590 ~ 1650 nm) / +10 ~ -70 dBm (1270 ~ 1650 nm) |
Total uncertainty (meter) |
0.3 dB (max) |
Resolution (meter) |
0.01 dB |
Polarization sensitivity (meter) |
< 0.05 dB (typical) |
ID tone detection & Autotest sensitivity (meter) |
Up to -50 dBm |
Autotest: |
1-way: compatible with other KI23400 & KI27400 models / 2-way: compatible with identical model only |
Tone detection (meter) |
150 ~ 9999 Hz ± 1 % |
USB interfaces |
micro-USB for power & software / USB-A Flash Drive |
Software |
KITS (free download) |
Power |
Battery / External power via micro-USB |
Battery life |
Laser/LED source: 80 hours (typical) in Autotest, Power Meter: 1000 hours (typical) |
Battery type |
2x AA (LR6) Alkaline / 2x AA NiMH AA (rechargeable via micro-USB) |
Auto off |
Selectable |
Display |
74 mm x 55 mm high contrast LCD |
Case |
Polycarbonate (body) / rubber (edges & corners), moisture resistance, 1-meter drop tested |
Memory |
For 8,000 fibers / Unlimited on USB Flash Drive |
Size |
190 mm x 105 mm x 35 mm |
Unit weight |
0.42 kg |
Shipping weight |
1.5 Kg |
Operating / storage temperature |
-15 to 55 °C / -25 to 70 °C |
Relative humidity |
0 ~ 95 % |
Warranty |
3 years |
Calibration validity (ISO/IEC 17025) |
3 years |